dc.description.abstractOBJECTIVE: Cochlear implants, while providing significant benefits to recipients, remain limited due to broad neural activation. Focussed multipolar stimulation (FMP) is an advanced stimulation strategy that uses multiple current sources to produce highly focussed patterns of neural excitation in order to overcome these shortcomings. APPROACH: This report presents single-source multipolar stimulation (SSMPsSSMPS), a novel form of stimulation based on a single current source and a passive current divider. Compared to conventional FMP with multiple current sources, SSMPS can be implemented as a modular addition to conventional (i.e. single) current source stimulation systems facilitating charge balance within the cochlea. As with FMP, SSMPS requires the determination of a transimpedance matrix to allow focusing of the stimulation. The first part of this study therefore investigated the effects of varying the probe stimulus (e.g. current level and pulse width) on the measurement of the transimpedance matrix. SSMPS was then studied using in vitro based measurements of voltages at non-stimulated electrodes along an electrode array in normal saline. The voltage reduction with reference to monopolar stimulation was compared to tripolar and common ground stimulation, two clinically established protocolsstimulation modes. Finally, a proof of principle in vivo test of SSMPS in a feline model was performed. MAIN RESULTS: A probe stimulus of at least 40 nC is required to reliably measure the transimpedance matrix. In vitro stimulation using SSMPS resulted in a significantly greater voltage reduction compared to monopolar, tripolar and common ground stimulation. Interestingly, matching measurement and stimulation parameters did not lead to an improved focussing performance. Compared to monopolar stimulation, SSMPS resulted in reduced spread of neural activity in the inferior colliculus, albeit with increased thresholds. SIGNIFICANCE: The present study demonstrates that SSMPS successfully limits the broadening of the excitatory field along the electrode array and a subsequent reduction in the spread of neural excitation.en_US
dc.description.sponsorshipAustralian Research Council through the linkage grant LP130100220 and the National Health and Medical Research Council (GNT1064375). The Bionics Institute acknowledges the support it receives from the Victorian Government through its Operational Infrastructure Support Program.en_US
dc.identifier.citationSenn, P., R. K. Shepherd, and J. B. Fallon. 2018. Focused electrical stimulation using a single current source. Journal of Neural Engineering. 15(5): 056018.en_US
dc.identifier.issn1741-2552
dc.identifier.urihttp://repository.bionicsinstitute.org:8080/handle/123456789/356
dc.language.isoenen_US
dc.publisherIOP Publishingen_US
dc.subjectCochlear implanten_US
dc.subjectCurrent focussingen_US
dc.subjectPhased arrayen_US
dc.subjectSpread of excitationen_US
dc.subjectElectrical stimulationen_US
dc.titleFocused electrical stimulation using a single current sourceen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
2018_Senn_Focusedelectrical.pdf
Size:
728.06 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections