Shepherd, R. K., P. Carter, Y. L. Enke, A. Thompson, B. Flynn, E. Trang, A. Dalrymple, and J. B. Fallon. 2020. Chronic intracochlear electrical stimulation at high charge densities: Reducing platinum dissolution

dc.contributor.authorShepherd, Robert
dc.contributor.authorCarter, Paul
dc.contributor.authorEnke, Ya Lang
dc.contributor.authorThompson, Alex
dc.contributor.authorFlynn, Brianna
dc.contributor.authorTrang, Ella
dc.contributor.authorDalrymple, Ashley
dc.contributor.authorFallon, James
dc.date.accessioned2020-11-23T05:11:58Z
dc.date.available2020-11-23T05:11:58Z
dc.date.issued2020-09
dc.description.abstractOBJECTIVE: Cochleae of long-term cochlear implant users have shown evidence of particulate platinum (Pt) corroded from the surface of Pt electrodes. The pathophysiological effect of Pt within the cochlea has not been extensively investigated. We previously evaluated the effects of Pt corrosion at high charge densities and reported negligible pathophysiological impact. The present study extends this work by examining techniques that may reduce Pt corrosion. APPROACH: Deafened guinea pigs were continuously stimulated for 28 days using biphasic current pulses at extreme charge densities using: (i) electrode shorting; (ii) electrode shorting with capacitive coupling (CC); or (iii) electrode shorting with alternating leading phase (AP). On completion of stimulation, cochleae were examined for corrosion product, tissue response, auditory nerve (AN) survival and trace levels of Pt; and electrodes examined for surface corrosion. MAIN RESULTS: Pt corrosion was evident at > 200 μC/cm2/phase; the amount dependent on charge density (p < 0.01) and charge recovery technique (p < 0.01); reduced corrosion was apparent using CC. Tissue response increased with charge density (p < 0.007); cochleae stimulated at > 200 μC/cm2/phase exhibited a vigorous response including a focal region of necrosis and macrophages. Notably, tissue response was not dependent on the charge recovery technique (p = 0.56). Despite stimulation at high charge densities resulting in significant levels of Pt corrosion, there was no stimulus induced loss of ANs. SIGNIFICANCE: Significant increases in tissue response and Pt corrosion were observed following stimulation at high charge densities. Charge recovery using CC, and to a lesser extent AP, reduced the amount of Pt corrosion but not the tissue response. Stimulation at change densities an order of magnitude higher than those used when programming cochlear implant recipients in the clinic, produced a vigorous tissue response and corrosion products without evidence of neural loss.en_US
dc.description.sponsorshipThis work was supported by National Institute on Deafness and Other Communications Disorders (R01DC015031) and Cochlear Ltd. The Bionics Institute acknowledges support of the Victorian Government through Operational Infrastructure Support Program.en_US
dc.identifier.citationShepherd, R. K., P. Carter, Y. L. Enke, A. Thompson, B. Flynn, E. Trang, A. Dalrymple, and J. B. Fallon. 2020. Chronic intracochlear electrical stimulation at high charge densities: Reducing platinum dissolution. Journal of Neural Engineering. 17: 056009.en_US
dc.identifier.issn1741-2552
dc.identifier.urihttp://repository.bionicsinstitute.org:8080/handle/123456789/410
dc.language.isoenen_US
dc.publisherIOP Publishingen_US
dc.subjectElectrical stimulationen_US
dc.subjectNeural prosthesisen_US
dc.subjectPlatinum electrodeen_US
dc.subjectStimulation safetyen_US
dc.subjectCorrosionen_US
dc.subjectTissue responseen_US
dc.titleShepherd, R. K., P. Carter, Y. L. Enke, A. Thompson, B. Flynn, E. Trang, A. Dalrymple, and J. B. Fallon. 2020. Chronic intracochlear electrical stimulation at high charge densities: Reducing platinum dissolutionen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
2020_Shepherd_Chronicintracochlear.pdf
Size:
56.67 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:
Collections