Cortical activation following chronic passive implantation of a wide-field suprachoroidal retinal prosthesis
Loading...
Date
2014-06
Journal Title
Journal ISSN
Volume Title
Publisher
IOP Publishing
Abstract
OBJECTIVE:
The research goal is to develop a wide-field retinal stimulating array for prosthetic vision. This study aimed at evaluating the efficacy of a suprachoroidal electrode array in evoking visual cortex activity after long term implantation.
APPROACH:
A planar silicone based electrode array (8 mm × 19 mm) was implanted into the suprachoroidal space in cats (ntotal = 10). It consisted of 20 platinum stimulating electrodes (600 μm diameter) and a trans-scleral cable terminated in a subcutaneous connector. Three months after implantation (nchronic = 6), or immediately after implantation (nacute = 4), an electrophysiological study was performed. Electrode total impedance was measured from voltage transients using 500 μs, 1 mA pulses. Electrically evoked potentials (EEPs) and multi-unit activity were recorded from the visual cortex in response to monopolar retinal stimulation. Dynamic range and cortical activation spread were calculated from the multi-unit recordings.
MAIN RESULTS:
The mean electrode total impedance in vivo following 3 months was 12.5 ± 0.3 kΩ. EEPs were recorded for 98% of the electrodes. The median evoked potential threshold was 150 nC (charge density 53 μC cm(-2)). The lowest stimulation thresholds were found proximal to the area centralis. Mean thresholds from multiunit activity were lower for chronic (181 ± 14 nC) compared to acute (322 ± 20 nC) electrodes (P < 0.001), but there was no difference in dynamic range or cortical activation spread.
SIGNIFICANCE:
Suprachoroidal stimulation threshold was lower in chronic than acute implantation and was within safe charge limits for platinum. Electrode-tissue impedance following chronic implantation was higher, indicating the need for sufficient compliance voltage (e.g. 12.8 V for mean impedance, threshold and dynamic range). The wide-field suprachoroidal array reliably activated the retina after chronic implantation.
Description
Keywords
blindness, retinal prosthesis, suprachoroidal, electrical stimulation
Citation
Villalobos, J., J. B. Fallon, D. A. X. Nayagam, M. N. Shivdasani, C. D. Luu, P. J. Allen, R. K. Shepherd and C. E. Williams (2014). Cortical activation following chronic passive implantation of a wide-field suprachoroidal retinal prosthesis. Journal of Neural Engineering 11(4): 046017