In vitro and In vivo comparison of the charge injection capacity of platinum macroelectrodes
Loading...
Date
2015-03
Journal Title
Journal ISSN
Volume Title
Publisher
IEEE
Abstract
Platinum (Pt) is the most commonly used metal for stimulating electrodes. This study aims to determine the amount of charge that can be delivered without causing irreversible electrochemical reactions (charge injection capacity, Q inj) of Pt macroelectrodes (geometric surface area >0.001 cm(2)) in vitro and in vivo using voltage transient measurements. Pt macroelectrodes were stimulated with biphasic charge-balanced cathodic-first constant-current pulses in phosphate buffered saline. Potential excursions were measured (versus Ag/AgCl electrode) and used to determine Qinj. The in vitro Qinj were compared to those measured in vivo following: acute and chronic implantation close to the retina; chronic intracochlear implantation; and acute subdural implantation, in the cat. Qinj increased with pulsewidth from 35 to 54 μC/cm(2) for respective pulse widths of 100 to 3200 μs per phase in vitro. Qinj was significantly less in vivo. There was no significant difference in Qinj between acutely (3.84 to 16.6 μC/cm(2) with pulsewidths of 100 to 3200 μs) and chronically (6.99 to 15.8 μC/cm(2) with pulsewidths of 200 to 3200 μs) implanted suprachoroidal electrodes. Intracochlear Qinj was not different to suprachoroidal Qinj, while subdural Qinj was significantly less than the suprachoroidal Q inj (p < 0.05). These results have important implications in providing guidelines on Qinj for the safe use of Pt stimulating macroelectrodes and question the relevance of measuring Qinj in vivo using voltage transients.
Description
Keywords
biomedical electrodes, blindness, electrical stimulation, electrochemical processes, implantable biomedical devices, in vitro, in vivo, neural prosthesis, visual prosthesis
Citation
Leung, R. T., Shivdasani, M. N., Nayagam, D. A. X., & Shepherd, R. K. (2015). In vitro and In vivo comparison of the charge injection capacity of platinum macroelectrodes. IEEE Trans Biomed Eng, 62(3), 849-85