Visual Cortex Responses to Single-and Simultaneous Multiple-Electrode Stimulation of the Retina: Implications for Retinal Prostheses

dc.contributor.authorShivdasani, Mohit
dc.contributor.authorFallon, James
dc.contributor.authorLuu, Chi
dc.contributor.authorCicione, Rosemary
dc.contributor.authorAllen, Penny
dc.contributor.authorMorley, John
dc.contributor.authorWilliams, Chris
dc.date.accessioned2016-03-04T05:28:42Z
dc.date.available2016-03-04T05:28:42Z
dc.date.issued2012
dc.description.abstractPurpose: The aim of this study was to compare simultaneous stimulation of multiple electrodes to single electrode stimulation in a retinal prosthesis. Methods: A platinum electrode array was implanted into the suprachoroidal space in six normally-sighted anesthetized cats. Multi-unit activity from the primary visual cortex in response to retinal stimulation was recorded. Cortical thresholds, yield of responses, dynamic ranges, and the spread of retinal activation were measured for three modes of stimulation; single electrode, half-row (6-electrode horizontal line) and column (7-electrode vertical line). Results: Stimulation of the best half-rows and columns was found to elicit activity with higher yield and lower charge thresholds per electrode compared to the best single electrodes. Dynamic ranges between the three modes were similar. As expected, peak voltages measured for columns and half-rows were lower than those measured for single electrodes. Spread of retinal activation, determined by the increase in threshold with distance in the retina from the best site was found to be similar between single and multiple electrode stimulation but dependent on orientation. Conclusions: The lower thresholds, higher yield, equivalent dynamic ranges and equivalent spread of retinal activation observed from simultaneous stimulation of multiple electrodes may be due to current and/or neural summation within the retina. Such stimulation techniques could be useful for the presentation of lines and edges of objects using a suprachoroidal retinal stimulator with low voltage compliance. Furthermore, the results suggest more complex visual processing strategies in addition to sequential stimulation of individual electrodes should be considered for retinal prostheses.en_US
dc.description.sponsorshipThis work was supported by the Australian Research Council through its Special Research Initiative in Bionic Vision Science and Technology awarded to Bionic Vision Australia and the Bertalli Family Foundation to the Bionics Institute. The Bionics Institute wishes to acknowledge the support it receives from the Victorian Government through its Operational Infrastructure Programen_US
dc.identifier.citationShivdasani, M. N., Fallon, J. B., Luu, C. D., Cicione, R., Allen, P. J., Morley, J. W., & Williams, C. E. (2012). Visual Cortex Responses to Single-and Simultaneous Multiple-Electrode Stimulation of the Retina: Implications for Retinal Prostheses. Investigative ophthalmology & visual science, 53(10), 6291-6300.en_US
dc.identifier.issn1552-5783
dc.identifier.urihttp://repository.bionicsinstitute.org:8080/handle/123456789/191
dc.language.isoenen_US
dc.publisherThe Association for Research in Vision and Ophthalmologyen_US
dc.relation.ispartofseries53;10
dc.titleVisual Cortex Responses to Single-and Simultaneous Multiple-Electrode Stimulation of the Retina: Implications for Retinal Prosthesesen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
2012 Shivdasani, Single and Simultaneous.pdf
Size:
911.47 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections