A Flexible Wireless System for Preclinical Evaluation of Retinal Prosthesis

dc.contributor.authorThien, Patrick
dc.contributor.authorMillard, Rodney
dc.contributor.authorEpp, Stephanie
dc.contributor.authorNayagam, David
dc.date.accessioned2018-02-12T00:38:12Z
dc.date.available2018-02-12T00:38:12Z
dc.date.issued2018-01
dc.description.abstractIn this paper, we present a novel stimulation controller and monitoring system for evaluating the safety and efficacy of implantable neuroprosthetic devices in a preclinical setting. It features a programmable controller designed to be worn in a custom backpack by freely moving feline subjects. A custom controller powered two, modified, 22-channel clinical stimulators simultaneously. The controller and stimulators together weighed 140 g and measured 85 × 70 × 35 mm3. Power was supplied from a 3350 mAh lithium-ion battery. A Bluetooth-enabled laptop-PC base station managed up to six systems and allowed the remote adjustment of the stimulation amplitude and automated collection of stimulator telemetry data. The initial application was for the chronic safety testing of a 44-channel retinal prosthesis. Thirteen feline subjects were implanted with a suprachoroidal electrode array, which was then stimulated or monitored continuously for an average of 54.5 d. Batteries were changed twice weekly and electrode impedances were recorded hourly. This allowed broken electrodes and other issues to be quickly identified and addressed. The ability to remotely control the stimulation amplitude minimised the amount of subject handling that was required, likely reducing subject stress. Existing preclinical evaluation systems are either designed for short-term experiments and have many features but limited battery life, or for long-term static stimulation and are long-lived, but with restricted stimulation parameters and channel counts. Here, we have described a system designed to improve the chronic safety testing of electrode arrays. While it was used here with a suprachoroidal retinal implant, it could be readily adapted for other preclinical models requiring continuous, deterministic stimulation over a prolonged period.en_US
dc.description.sponsorshipThis research was supported by the Australian Research Council (ARC) through its Special Research Initiative (SRI) in Bionic Vision Science and Technology grant to Bionic Vision Australia (BVA). Further support was provided by a National Health and Medical Research Council Development Grant (APP1120664). The Bionics Institute receives Operational Infrastructure Support from the Victorian Government. The Bionics Institute would also like to acknowledge support from the Bertalli Family Trust and the J T Reid Charitable Trust.en_US
dc.identifier.citationThien, P. C., R. Millard, S. B. Epp, and D. A .X. Nayagam. 2018. A Flexible Wireless System for Preclinical Evaluation of Retinal Prosthesis. Sensors and Materials. 30(2): 269 - 86.en_US
dc.identifier.issn0914-4935
dc.identifier.urihttp://repository.bionicsinstitute.org:8080/handle/123456789/282
dc.language.isoenen_US
dc.publisherMYU KKen_US
dc.subjectElectrical stimulationen_US
dc.subjectWireless monitoringen_US
dc.subjectFreely moving animalsen_US
dc.subjectRetinal prosthesisen_US
dc.subjectPreclinicalen_US
dc.titleA Flexible Wireless System for Preclinical Evaluation of Retinal Prosthesisen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
2018_Thien_Aflexible.pdf
Size:
4.38 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:
Collections